6010 SERIES
ENVIRONMENTAL CHAMBER
OPERATIONS MANUAL
FOR MODELS
6010
6011
6012
6013
6015
6017
PO Box 715
Marietta, OH 45750
800-648-3042 740-373-6809
Fax 740-374-3760
www.caronproducts.com
service@caronproducts.com
Page view 0
1 2 3 4 5 6 ... 67 68

Summary of Contents

Page 1

6010 SERIES ENVIRONMENTAL CHAMBER OPERATIONS MANUAL FOR MODELS 6010 6011 6012 6013 6015 6017 PO Box 715 Marietta, OH 45750 800-648-3

Page 2

6010 Series Operations Manual Rev E 2/14/2012 Page 10 of 68 SECTION 2 – EQUIPMENT OVERVIEW Congratulations! You have just purchased

Page 3 - TABLE OF CONTENTS

6010 Series Operations Manual Rev E 2/14/2012 Page 11 of 68 SECTION 2 – EQUIPMENT OVERVIEW -- CONTINUED Temperature Controller Humidity

Page 4

6010 Series Operations Manual Rev E 2/14/2012 Page 12 of 68 SECTION 3 -- INSTALLATION Unpacking Your new unit has been thoroughly packaged

Page 5 - EQUIPMENT LIMITED WARRANTY

6010 Series Operations Manual Rev E 2/14/2012 Page 13 of 68 The unit requires a dedicated electrical connection. Power requirements vary de

Page 6

6010 Series Operations Manual Rev E 2/14/2012 Page 14 of 68 Stacking two Units The CARON 6010 family is designed to allow two units

Page 7

6010 Series Operations Manual Rev E 2/14/2012 Page 15 of 68 Use only distilled or deionized water with a resistivity between 50K-CM and 1M

Page 8

6010 Series Operations Manual Rev E 2/14/2012 Page 16 of 68 Connect each chamber to a grounded circuit. Failure to do so could result in el

Page 9 - WARNINGS

6010 Series Operations Manual Rev E 2/14/2012 Page 17 of 68 SECTION 4 – ACCESSORY INSTALLATION Connecting Alarm Contacts (ALRM301) With t

Page 10

6010 Series Operations Manual Rev E 2/14/2012 Page 18 of 68 Installing Carboy Water System (BOTL301) Models 6010, 6015 & 6017 can be pu

Page 11

6010 Series Operations Manual Rev E 2/14/2012 Page 19 of 68 Connecting Communications or Analog Outputs (DLUX301 & DLUX302) With the

Page 12 - SECTION 3 -- INSTALLATION

6010 Series Operations Manual Rev E 2/14/2012 Page 2 of 68 Dear Valued Customer: Thank you for purchasing CARON Products & Services

Page 13

6010 Series Operations Manual Rev E 2/14/2012 Page 20 of 68 RS-485 communications are intended to communicate with a PC using ModBus RTU or

Page 14

6010 Series Operations Manual Rev E 2/14/2012 Page 21 of 68 Installing the Heatless Dryer Package (DRYR301) The Heatless Dryer Package

Page 15

6010 Series Operations Manual Rev E 2/14/2012 Page 22 of 68 Incoming compressed air must be free of dirt, oil, moisture and filtered to 10 m

Page 16

6010 Series Operations Manual Rev E 2/14/2012 Page 23 of 68 3. Attach 3/8” orange tubing from the chamber to the flow meter outlet.

Page 17

6010 Series Operations Manual Rev E 2/14/2012 Page 24 of 68 Lights should only be used in a non-condensing environment. Connecting the Fluor

Page 18

6010 Series Operations Manual Rev E 2/14/2012 Page 25 of 68 Installing Drain Water Pump (PUMP301) In applications where a floor drain is no

Page 19

6010 Series Operations Manual Rev E 2/14/2012 Page 26 of 68 Installing the Stacking Kit (STCK301) Two environmental chambers can be st

Page 20

6010 Series Operations Manual Rev E 2/14/2012 Page 27 of 68 SECTION 5 – OPERATION With the chamber properly installed and the appropriat

Page 21

6010 Series Operations Manual Rev E 2/14/2012 Page 28 of 68 Changing the Humidity Set-point The standard cabinet has a humidity set-poin

Page 22

6010 Series Operations Manual Rev E 2/14/2012 Page 29 of 68 SECTION 6 – ACCESSORY OPERATION Using the Carboy Water System (BOTL301) To fi

Page 23 - Flow Meter

6010 Series Operations Manual Rev E 2/14/2012 Page 3 of 68 TABLE OF CONTENTS Section 1 – Warranty ...

Page 24 - Light Bank

6010 Series Operations Manual Rev E 2/14/2012 Page 30 of 68 Operation of the Defrost System (DEFR301) A Defrost System is used to mai

Page 25

6010 Series Operations Manual Rev E 2/14/2012 Page 31 of 68 Operation of the Delux Controller System (DLUX301 & DLUX302) The chambers

Page 26

6010 Series Operations Manual Rev E 2/14/2012 Page 32 of 68 Analog Outputs: With DLUX301 & DLUX302 controllers, there is an analog outp

Page 27 - SECTION 5 – OPERATION

6010 Series Operations Manual Rev E 2/14/2012 Page 33 of 68 Set the Scale Low value to correspond with the minimum value of the

Page 28

6010 Series Operations Manual Rev E 2/14/2012 Page 34 of 68 Ramp & Soak: A ramp and soak control system is included with DLUX301 &

Page 29

6010 Series Operations Manual Rev E 2/14/2012 Page 35 of 68 Operation of the Heatless Dryer Package (DRYR301) The flow meter shows the flow

Page 30

6010 Series Operations Manual Rev E 2/14/2012 Page 36 of 68 Controlling the Diurnal Lights (6012) and Optional Fluorescent Lights (LGHT303)

Page 31

6010 Series Operations Manual Rev E 2/14/2012 Page 37 of 68 Operating the timer The power switch enables and disables the lights and c

Page 32

6010 Series Operations Manual Rev E 2/14/2012 Page 38 of 68 Examples for how to start your timer during day or night modes Night Start Pro

Page 33

6010 Series Operations Manual Rev E 2/14/2012 Page 39 of 68 Accumulate or Reset Factory default is set so the timer continues where

Page 34

6010 Series Operations Manual Rev E 2/14/2012 Page 4 of 68 Section 7 – Calibration ...

Page 35

6010 Series Operations Manual Rev E 2/14/2012 Page 40 of 68 Controlling the LED Lighting (LGHT304) The LED lighting system creates lo

Page 36

6010 Series Operations Manual Rev E 2/14/2012 Page 41 of 68 Operating the timer The power switch enables and disables the lights and c

Page 37

6010 Series Operations Manual Rev E 2/14/2012 Page 42 of 68 Accumulate or Reset Factory default is set so the timer continues where

Page 38

6010 Series Operations Manual Rev E 2/14/2012 Page 43 of 68 Interior Electrical Outlet (OUTL301) An optional interior electrical out

Page 39 - Accumulate

6010 Series Operations Manual Rev E 2/14/2012 Page 44 of 68 Operation of Front Mounted 6” Recorders (RCDR307, RCDR308) Built in 6” ink pen

Page 40

6010 Series Operations Manual Rev E 2/14/2012 Page 45 of 68 and metal pen arm be removed as an assembly. Unsnap the plastic "U" cl

Page 41

6010 Series Operations Manual Rev E 2/14/2012 Page 46 of 68 If a calibration adjustment is required for a single pen recorder, use the left

Page 42

6010 Series Operations Manual Rev E 2/14/2012 Page 47 of 68 Operation of Front Mounted 10” Recorders (RCDR309, RCDR310) Built in 10” th

Page 43

6010 Series Operations Manual Rev E 2/14/2012 Page 48 of 68 move off of the chart allowing you to place the recording chart paper onto the r

Page 44

6010 Series Operations Manual Rev E 2/14/2012 Page 49 of 68 Recorder Calibration: If calibration is required for single input recorders, us

Page 45

6010 Series Operations Manual Rev E 2/14/2012 Page 5 of 68 SECTION 1- WARRANTY INFORMATION EQUIPMENT LIMITED WARRANTY Please review this s

Page 46 - use the left (#1) and right

6010 Series Operations Manual Rev E 2/14/2012 Page 50 of 68 Operation of Side Mounted Recorders (RCDR311, RCDR312) Side mounted Honeywell

Page 47

6010 Series Operations Manual Rev E 2/14/2012 Page 51 of 68 4. Slip the new chart under the pen lifter, pens and time index; and press the

Page 48

6010 Series Operations Manual Rev E 2/14/2012 Page 52 of 68 SECTION 7 – CALIBRATION The temperature and humidity systems can all be

Page 49

6010 Series Operations Manual Rev E 2/14/2012 Page 53 of 68 Calibrating the Humidity If humidity calibration is needed, the following steps

Page 50

6010 Series Operations Manual Rev E 2/14/2012 Page 54 of 68 SECTION 8 – ALARMS Alarm System Overview The chamber control system is equippe

Page 51

6010 Series Operations Manual Rev E 2/14/2012 Page 55 of 68 SECTION 9 – PREVENTATIVE MAINTENANCE The CARON chamber has been robustly design

Page 52 - SECTION 7 – CALIBRATION

6010 Series Operations Manual Rev E 2/14/2012 Page 56 of 68 SECTION 10 – SPECIFICATIONS MODEL 6010/6015 6011 6012 6013 6017 Temperature Ran

Page 53 - Infinity Key

6010 Series Operations Manual Rev E 2/14/2012 Page 57 of 68 This graph illustrates the operating temperature & humidity range of models

Page 54 - SECTION 8 – ALARMS

6010 Series Operations Manual Rev E 2/14/2012 Page 58 of 68 SECTION 11 – ELECTRICAL SCHEMATICS

Page 55

6010 Series Operations Manual Rev E 2/14/2012 Page 59 of 68

Page 56 - SECTION 10 – SPECIFICATIONS

6010 Series Operations Manual Rev E 2/14/2012 Page 6 of 68 This writing is a final and complete integration of the agreement between CARON a

Page 57

6010 Series Operations Manual Rev E 2/14/2012 Page 60 of 68 SECTION 12 – TROUBLESHOOTING Problem -- Unit will not turn on  Is the unit

Page 58

6010 Series Operations Manual Rev E 2/14/2012 Page 61 of 68 SECTION 13 – SPARE / REPLACEMENT PARTS General Part Number Description MTR-13

Page 59

6010 Series Operations Manual Rev E 2/14/2012 Page 62 of 68 Options Related Part Number Description Option LGT-156 24W HO florescent bulb

Page 60

6010 Series Operations Manual Rev E 2/14/2012 Page 63 of 68 The temperature and humidity controllers are factory programmed for precise cont

Page 61

6010 Series Operations Manual Rev E 2/14/2012 Page 64 of 68 APPENDIX A – RAMP & SOAK EXAMPLE General An efficient way to program

Page 62 - Options Related

6010 Series Operations Manual Rev E 2/14/2012 Page 65 of 68 8. Select the step type by pressing the ADVANCE button. 9. Use the UP and

Page 63

6010 Series Operations Manual Rev E 2/14/2012 Page 66 of 68 Event Output 1 Ent1 oFF Event Output 2 Ent2 oFF Step 2 P1 2 Soak step type S.

Page 64

6010 Series Operations Manual Rev E 2/14/2012 Page 67 of 68 End step type S.tyP End Type of profile end (hold 30ºC) End HoLd Step 8 P1 8

Page 65

6010 Series Operations Manual Rev E 2/14/2012 Page 68 of 68 Caron Products and Services, Inc. 27640 State Route 7 Marietta, OH 45750

Page 66

6010 Series Operations Manual Rev E 2/14/2012 Page 7 of 68 EQUIPMENT INTERNATIONAL LIMITED WARRANTY Please review this section before reque

Page 67

6010 Series Operations Manual Rev E 2/14/2012 Page 8 of 68 This writing is a final and complete integration of the agreement between CARON a

Page 68 - DECLARATION OF CONFORMITY

6010 Series Operations Manual Rev E 2/14/2012 Page 9 of 68 INTERNATIONAL SYMBOLS AND DEFINITIONS WARNINGS